Peptide purification improvements with flash column chromatography by modulating mobile phase pH

Peptides, by nature, are composed of amino acids with potentially ionizable chemical moieties. The ionization state of any of these moieties can significantly impact the peptide’s chromatographic behavior, both in terms of peak shape and retention by the solid support.  Peptide purification by reversed-phase chromatography, however, almost exclusively includes an acidic additive to the mobile phase solvents, maintaining the solution at a pH of 2-3 throughout the purification cycle.  But have you ever considered trying an alternative additive in the mobile phase to improve your purification results?

In the following post I discuss the impact of mobile phase pH in the purification of oxytocin (CYIQNCPLG-NH2), a 9-amino acid peptide that requires disulfide bond-mediated cyclization for its biological activity.

In the case of oxytocin, it is critical to separate the linear from cyclized peptide if any biological assays are to be completed in the future.  As with all other peptides I’ve synthesized, I analyzed the crude peptide mixtures by analytical HPLC-MS containing formic acid in the mobile phase solvents.  The addition of 0.1% formic acid attains a pH of 2.5-2.7 under these conditions.  I was pleased to see during this analysis that the linear and cyclized peptides are well resolved, Figure 1.

crude-analytical
Figure 1: Crude analytical HPLC of a peptide sample containing both linear and disulfide-mediated cyclized oxytocin.

Unfortunately though, after many attempts to separate these two peptides by varying the gradient with a Biotage® SNAP Ultra C18 cartridge, I resigned myself to the fact that flash column chromatography is not selective enough to separate the linear from cyclized oxytocin peptides, Figure 2.

optimized-oxytocin-ph-2
Figure 2: Optimized flash column chromatography gradient to separate linear from cyclized oxytocin dissolved in DMSO. The gradient runs from 15% to 30% acetonitrile with mobile phase solvents containing 0.1% TFA.

Rather than give up on flash column chromatography and return to HPLC (I only have an analytical system with no fraction collector), I thought about the sequence of oxytocin and any features that may chromatographically differentiate the linear from cyclized peptide.  Glaring back at me were two reduced Cys sidechains (pKa about 8.37) that could be ionized under basic conditions.  After some searching for mobile phase additives that would achieve pH>9 but still maintain compatiblity with LCMS, I settled on 0.1% ammonium hydroxide (pH ~ 10.7).  Ideally, the pH should be 1-2 units above the pKa to ensure complete deprotonation of the Cys sulfur atoms and quality chromatography.

I was shocked by the dramatic difference in chromatography I observed after injecting oxytocin, dissolved in DMSO, with the basic mobile phase, Figure 3.  Please note that this sample was monitored with negative ionization mode on the mass spectrometer.

optimized-oxytocin-ph10
Figure 3: Optimized flash column chromatography gradient to separate linear from cyclized oxytocin dissolved in DMSO. The gradient runs from 5% to 30% acetonitrile with mobile phase solvents containing 0.1% ammonium hydroxide.

I analyzed fractions within each of the three major peaks by analytical HPLC-MS and found that the linear peptide was exclusively contained within the peak 1 and that the cyclized peptide was contained in both peaks 2 and 3 (I checked, nothing with peptidic character is present in the large peak at the end of the run).  I suspect that the cyclized peptide is resolved as two peaks due to partial deprotonation of the sidechain as the peptide is dissolved in DMSO.

To confirm that the DMSO was in fact influencing the accessibility of the peptide solution to the mobile phase and the pH modifiers, I dissolved the peptide mixture in 5% aq. acetonitrile containing 0.1% ammonium hydroxide.  I wanted to ensure the peptides were fully deprotonated prior to injecting onto the SNAP Ultra C18 cartridge, Figure 4.

mobile-phase-solvation
Figure 4: Optimized flash column chromatography gradient to separate linear from cyclized oxytocin dissolved in 5% aqueous acetonitrile containing 0.1% ammonium hydroxide. The gradient runs from 5% to 30% acetonitrile with mobile phase solvents containing 0.1% ammonium hydroxide.

Gratifyingly, the general peak separation pattern is maintained when the peptide mixture is dissolved in the initial mobile phase conditions.  I should mention that the 20 mg crude sample was not fully solubilized in the 4-mL liquid solution prior to injecting onto the SNAP Ultra C18 cartridge.  This is the likely cause for some of the observed peak shape decomposition.  Interestingly, the linear peptide is not retained by the the column in the absence of DMSO.  However, the cyclized peptide is still clearly resolved into two peaks, peaks 2 and 3 respectively.  As the peptide mixture is dissolved in the basic mobile phase, I conclude that the two peaks arise from the partial deprotonation of the Tyr hydroxyl group (pKa ~10.46).  The additional negative charge on such a small peptide would absolutely influence it’s retention by the C18 column, thereby yielding the observed peak distribution.

Although the resolution of flash column chromatography is less superior than HPLC, I have demonstrated here that with a little thought, a peptide can be purified from it’s contaminants by modulating the pH of the mobile phase.  I will certainly be trying this technique for longer peptides in the future just to avoid the HPLC!  Have you ever tried purifying peptides with alternative pH mobile phase additives?

 

Published by

Elizabeth Denton

Elizabeth Denton is the Peptide Application specialist for Biotage and handles all things peptide synthesis and workflow related, in addition to troubleshooting problematic syntheses. Elizabeth recently began her career at Biotage after working as a peptide scientist for a small biotech company. Elizabeth earned a bachelors degree in Chemistry from Arizona State University and a Ph. D. in Chemistry from Yale University with a focus in Chemical Biology. Throughout her thesis work, Elizabeth focused on projects in which peptides, containing both natural and unnatural amino acids, were designed to target and activate natural proteins.

4 thoughts on “Peptide purification improvements with flash column chromatography by modulating mobile phase pH”

  1. I was wondering if using such base is actually compatible with the columns you were using !!
    did you notice any cracking or non-uniform flow ?

    1. Hello Ahmed,
      Thank you for comment. Column stability to pH is something to always be aware of, even with HPLC columns. The C18 functionalized silica packed in the cartridge I used for these experiments was evaluated for stability at pH 10 with constant flow for >24 hours with no loss of C18 observed. After the injections though, I washed the column with neat acetonitrile for storage. I did not observe any cracking or non uniform flow during the injections, nor in subsequent injections. I have used the same cartridge since these purifications with acidified mobile phase and have observed no loss in resolution.

  2. Beth, good work here. I think others will find this use of pH adjustment quite helpful. I’m curious to see what other replies you receive. Thanks for the effort!

  3. Nicely done, thank you for sharing your experience. It is something I may try in the future when I get in a purification pinch!