How do I develop a reversed-phase flash column chromatography method?

Method transfer from reversed phase TLC (thin layer chromatography) to reversed phase flash column chromatography can be very challenging.  Because of this, I often recommend using HPLC for reversed phase flash chromatography method development. This really is a straight-forward process if you start with the right HPLC column and know a little about your HPLC system’s detector.

In this post I will share some tips on how to develop a reversed-phase flash purification method using HPLC.

In a previous post I discussed the challenges that reversed phase TLC poses for flash column purification, the primary issue being that surface wetting of the C18 stationary phase provides inconsistent prediction of chromatography conditions.

So, with reversed phase TLC not being a reliable method development tool for flash column chromatography what other options are available?  Well, I normally recommend reversed-phase HPLC with a similar bonded phase chemistry, e.g. C18.

Method development using HPLC

Now that you know the reasoning behind why method development on HPLC for reversed-phase flash chromatography is preferred, I need to explain how to do the conversion.

Simply put, the process is to create a gradient on an HPLC system and then to duplicate it on a flash system. It is best to use a HPLC column that is packed with the same media as the flash column (Biotage offers HPLC columns packed with their flash C18 media, they are known as scaling columns). Using a scaling column for method development eliminates any separation quality differences related to particle size and the resulting resolution/efficiency impact on the separation.

Using an off-the-shelf, commercially available HPLC column will certainly point you in the right direction regarding the gradient method but due to media particle size and selectivity differences, the results may not translate as well to flash as you might want.

To illustrate the challenges associated with using a generic reversed-phase HPLC column for method development, I conducted an experiment where I used a commercially available HPLC column to develop a suitable separation method and then transferred the exact method to flash chromatography.

Table 1 shows the HPLC gradient, which was based on time, in tabular form. In Figure 1 both the gradient and results are shown in graphical form.  For the HPLC a Phenomenex Gemini® (4.6 x 150 mm, 5 µm) was used and found to provide adequate separation results for transferring the method to reversed-phase flash chromatography.

Table 1. Final HPLC gradient

Time% MethanolFlow rate (mL/min)
0551
4551
24901
26901
26.011001
321001
Reversed-phase HPLC gradient and resulting HPLC trace used for flash chromatography method development.
Figure 1. Reversed-phase HPLC gradient (top) and resulting HPLC trace used for flash chromatography method development. The chromatogram was sized to match the gradient profile to help indicate where in the gradient each compound elutes. The column was a 4.6 x 150 mm Phenomenex Gemini 5 um C18 operated at a flow rate of 1 mL/min. The solvents were water and methanol and the load amount was 1.667 ug.

To accomplish this conversion I made note of the HPLC retention time for peak 1 (1.82 min) which indicates the void time.  Since the flow rate was 1 mL/min that means the void volume for the column is 1.82 mL which is equal to 1 column volume (CV).  Converting the HPLC method from time to column volumes eliminates the need to know column size and flow rate when duplicating it on a flash system; Table 2 displays the gradient in Table 1 in terms of column volumes.

Table 2. HPLC method converted from time to column volumes

CV% MethanolFlow rate (mL/min) 
0551
2551
12901
13901
13.011001
161001

I programmed the flash system (I used an Isolera One™ system) with the CV-based gradient and selected a 12 gram SNAP Ultra C18™ (Biotage) to perform the purification, Figure 2.

Reversed phase flash method based on HPLC gradient
Figure 2. Reversed phase flash method based on HPLC gradient.

For the flash purification I loaded 100 mg of the sample. The results show a full separation between each compound but with some slightly different selectivity compared to the HPLC column.  In fact, the flash cartridge actually separates the third eluting peak (yellow) better than the HPLC column while the separation between peaks 4 and 5 is less than the HPLC, Figure 3.

So, what do these results really tell us? Well, the results indicate that the generic reversed-phase HPLC column was able to approximate the potential flash cartridge performance but it was not a direct correlation. This is why the use of a scaling column packed with the same media as the flash cartridge makes sense – the selectivity and resolving power will be identical making scale-up to flash much more straight-forward.

Reversed-phase purification using the HPLC gradient
Figure 3. Flash reversed-phase purification using the HPLC-developed gradient.  The gradient segments are matched in terms of column volumes and the results are similar but not identical.  The flash cartridge is a 12 gram SNAP Ultra C18 operated at 20 mL/min with a sample load of 100 mg.

Scaling column based method development

To start, set some basic parameters then modify the gradient start and end points as needed to obtain a suitable separation. In Table 3, below, I show some example starting points for two Biotage C18 media cartridges.

Table 3. Suggested reversed-phase method development starting points

 KP-C18-HS scaling columnUltra C18 scaling column
Column size (mm x mm)4.6 x 2504.6 x 250
Particle size (um)5025
Flow rate (mL/min)11
Linear velocity (cm/min)7.56.7
DetectionDiode arrayDiode array
Run time, gradient (min)4349
Solvent AWaterWater
Solvent BMethanol or acetonitrileMethanol or acetonitrile
Equilibration10% B for 10 min. (3CV)10% B for 11.2 min. (3CV)
Gradient segment 110% B for 3.3 min. (1 CV)10% B for 3.7 min. (1 CV)
Gradient segment 210% to 100% B in 33.2 min (10 CV)10% to 100% B in 37.4 min (10 CV)
Gradient segment 3100% B for 6.6 min. (2 CV)100% B for 7.5 min. (2 CV)
Sample size (mg)less than 1less than 1

Basically, create a 3-segment linear gradient based on column volumes using the scaling column.  The scaling column’s column volume is the void time (t0) multiplied by the flow rate.

Equilibration     10% MeOH (or MeCN) for 3 CV

Segment 1           10% MeOH (or MeCN) for 1 CV

Segment 2           10% MeOH (or MeCN) to 100% in 10 CV

Segment 3           100% MeOH (or MeCN) for 2 CV

If your compounds elute too early, reduce the end % B to 50% and repeat (do not change the run time or flow rates).

If your compounds elute late in the gradient, increase the start % B to 50% and repeat (do not change the run time or flow rates).

Continue this process until you have an acceptable separation then follow steps 1 through 6.

Challenge the column’s loading capacity until your target compound has a resolution (Rs) ~1 from its nearest neighbors, this will be your loading limit which can be scaled up to any size flash cartridge. Most HPLC systems have software that will calculate resolution.

  1. Find out what the HPLC system’s flow cell path length is and divide it by the path length of your flash system’s detector (0.3 mm on all Isolera systems). This will help compensate your flash loading for the more sensitive analytical HPLC detection.
  2. Multiply your scaling column load by the number calculated above to give you an estimated flash system load for the scaling column.
  3. To choose the proper cartridge size for the amount of material you need to purify, check Table 4.
  4. On your flash system create the same method in CV choosing the appropriate cartridge media and size for your purification.
  5. Set your flash flow rate to match the scaling column linear velocity using Table 5 below.

Table 4. Scale up factors for reversed-phase*

Column size Scale factor
Scaling column 1
12 gram 4.6
30 gram 11.5
60 gram 23
120 gram 46
400 gram 153

*for more information on scaling factors I suggest you visit http://www.biotage.com/product-page/biotage-snap-ultra-c18. You will have to register on the website, but they you can download the Flash Purification Consumables brochure and review the very useful scale-up table on page 29.

Table 5. Flash column flow rates with equivalent linear velocities

Column size KP-C18-HS flow rate (mL/min) Ultra C18 flow rate (mL/min)
Scaling column 0.5 0.5
12 gram 10 12
30 gram 17 24
60 gram 30 40
120 gram 30 40
400 gram 105 135

If you follow this process you should find success moving a reversed-phase HPLC method to flash.

Share your reversed-phase flash method development techniques with the rest of the readers and suggest other topics you would like to see discussed.

Published by

Bob Bickler

Technical Specialist, Biotage

8 thoughts on “How do I develop a reversed-phase flash column chromatography method?”

    1. Hi Brett,

      Thank you for your question regarding C18 cartridge reusability. Yes, reversed-phase columns are reusable though they should never be stored in an acidic, basic, or buffered environment. I suggest storing in 80% or greater organic solvent (acetonitrile or methanol) which can often be used to clean lipophilic debris from the top of the cartridge.

      Bob

    1. Hi Dr. Patel,

      Thank you for your kind words. My goal for this blog is to provide helpful tips to chemists who use flash chromatography. Do you have any topics which you would like addressed in the blog?

      Best regards,

      Bob

    1. Hi Chris,

      I am pleased you found this helpful, that of course is the blog’s goal. Are there other topics that you would like addressed through the blog?

      Best regards,

      Bob

  1. Dear Bob,

    Thanks for a good article sharing. A lot of HPLC mobile phase was added additive such as phosphate buffer, acetic buffer or formic acid for better separation and resolution. May I know whether if the mobile phase for reverse phase HPLC using buffer or formic acid solution, can we direct convert to reverse phase flash purification using HPLC gradient method? Kindly advise.

    Thanks.

    Regards,
    Chew

    1. Hi Won Yin,

      Thank you for your question on mobile phase additives. As you mention additives to mobile phases are often used to enhance peak shape and alter retention times. In analytical HPLC inorganic buffers are often used to achieve these goals but in preparative HPLC and flash the use of volatile buffers and pH modifiers is strongly recommended because they simplify compound recovery post purification. Because of their ease of evaporation additives such as formic acid, ammonia, and their salts are commonly used. So, if you are developing a reversed-phase method that requires a pH modifier I suggest using volatile additives and volatile buffers.

      Best regards,

      Bob